DM5001 Controller/HB series Dynamometers
Torque dynamometer with a reliable hysteresis brake
Better performance, better usability, and upgraded functionality
Motor Testing Equipment(PDF: 4.6MB)
Torque Dynamometer (PDF:3.3MB)
(∗1) Select "WT1800E" in the “WT Compatible Command Type Setting” of WT5000. Up to six elements can be used.
In the point measurement mode based on our own control method, the time taken for the motor speed to stabilize (the “wait time”) can be set to eliminate the effect of the moment of inertia on the load torque value and rotation speed. And in the sequence measurement mode used for dynamic measurements, it is possible to minimize the effect of inertia on measurements by using correction values obtained through simple operations.
Setting up the conventional dynamometers involves adjusting a different set of P, I and D values for each motor being measured, which requires a knowledgeable and experienced operator. In Sugawara’s unique torque control method, there is no need for PID parameter setting. To obtain unambiguous data, all you have to do is fix the motor and shaft correctly and set up the points where torque or rotation speed measurements are to be made. The results will be the same no matter who performs the measurements.
Four measurement modes can be selected to suit each application.
You can easily measure the S-T (rotational speed-torque) characteristics simply by setting the measurement points. Torque control, rotational speed control and brake control can be combined in one measurement. The data is not affected by moment of inertia because it measures the value at a stable operating point.
By controlling the torque, rotation speed and braking over time, it is possible to perform various tests such as sweep measurements, endurance tests and cycle tests. This makes it possible to perform load simulations based on actual use conditions, and durability tests that combine test conditions and idling conditions and repeat it up to 10,000 cycles.
Pass/fail testing can be performed by setting up to five upper and lower limits for measurement items based on torque, rotation speed or braking. By specifying a file name in advance, you can automatically generate a file listing the time, sequence number and results of each measurement.
In this mode, you can set and control the torque, rotation speed or braking at a single point, and display all the corresponding measured values in real time. This mode can be used for simple behavior observations and for the measurement of starting torque.
The addition of extra I/O options provides you with various extended functions.
The graph displays 5 axes × 3 items in real time. We have improved the operability of this display by adding a graph axis auto setting function. You can also customize the graph line style and dot appearance.
It is also easy to change the power analyzer settings in software.
The dedicated TORQuick software is written in LabVIEW and can be flexibly adapted to diverse requirements. Please contact our sales department for further details.
We have improved the torque measurement accuracy of the hysteresis brake to ±0.1%. A total of 13 models are available, with rated torques ranging from 5 mN·m to 50 N·m. The three models with maximum torques of 50 mN·m, 100 mN·m and 200 mN·m can attain a world-class no-load rotation speed of 60,000 rpm due to our high-precision processing.
DM5001 is compatible with the following power meters for voltage and current measurements.
Yokogawa Test & Measurement Corporation
WT5000 (∗1)
WT1800E Series
WT310/WT330/WT300E series
HIOKI E.E. CORPORATION
PW3390/PW3335/PW3336/PW3337
You can choose a suitable one according to your application such as single-phase AC/DC motors and inverter motors.
Torque measurement | Sampling rate:1 ms (16 bit AD converter) Averaging:100 ms moving average Torque accuracy:±0.1% of full scale (accuracy of the system including DM5001 Controller, after calibration) |
---|---|
Rotation speed accuracy | ±0.01% ±2digit |
Load control parameter | Torque control:Servo-control of torque value Speed control:Servo-control of rotation speed Brake control:Open-loop control of brake current |
Output power | T[N・m]×S[r/min]×0.1047 |
Voltage, current, power, efficiency, power factor, and frequency | Measured by a connected power meter |
Power meter |
Yokogawa Test & Measurement Corporation WT5000 (∗1) WT1800/WT1800E series WT310/WT330/WT300E series HIOKI E.E. CORPORATION PW3390/PW3335/PW3336/PW3337 |
Analog output | Torque analog output:±10 VDC/Torque range full scale ±1% Speed analog output:±10 VDC/Speed range full scale ±1% |
Options of DM5001 | I/O option Digital input 4ch, Digital output 4ch Analog input 4ch, Analog output 2ch Contact output for power supply control (NO, NC)1ch |
Controller | DM5001 |
---|---|
Power | 100-240 VAC, 50/60 Hz |
Power consumption | 30 VA |
Temperatures | 0-40 ℃ |
Humidity | 20-90 %RH, without dew condensation |
Dimensions | W × H × D 430 × 148 × 360 mm |
Weight | 8 kg |
(∗1) Select "WT1800E" in the “WT Compatible Command Type Setting” of WT5000. Up to six elements can be used.